1. Significant improvement in furnace performance (rapid heating/rapid cooling):
Rapid heating and cooling is achieved by employing a ceramic heater with low thermal mass. This excellent heater features a wide temperature range: from ambient to 1050 ºC.
2. High performance and high reliability (pyrogram, inertness, analysis of high boiling point compounds):
The performance is guaranteed to meet the stated specifications such as the thermogram of evolved gas analysis and the pyrogram of instantaneous pyrolysis analysis.
3. Any sample form for analysis:
The evolved gas analysis, thermal desorption, flash pyrolysis, double-shot, and heart-cut EGA-GC/MS methods can be used to characterize polymer samples from multiple viewpoints.
【Mechanism】 * Move to Vimeo site by clicking the following link.
・ Operating mechanism of EGA (Evolved Gas Analysis) (Video)
・ Operating mechanism of Single-Shot Analysis (Video)
・ Operating mechanism of Double-Shot Analysis (Video)
Furnace temperature control range/temperature stability:
room temperature +10 to 1050 ºC (1 ºC increment) / within ± 0.1 ºC
Heating furnace heating rate:
Max. 600 ºC / min (every 1 ºC / min)
Interface temperature control range/temperature stability:
40 to 450 ºC (1 ºC increment) / within ±0.1 ºC
Performance guarantee (polystyrene pyrogram with MS detector):
Reproducibility of the trimer and internal standard ratio is 2 % or less (RSD)
Performance guarantee (polystyrene thermogram with MS detector):
Peak top temperature reproducibility is 0.3 % or less (RSD)
Requirements:
gas chromatograph (with split/splitless inlet), etc.
Size (heating furnace):
76 (W) x 143 (D) x 215 (H) mm, 1.6 kg
Required power:
100-115 VAC (400 W max)